IEC 60601-1-2 ED. 4.1 New requirement, immunity to proximity magnetic fields based on IEC 61000-4-39

Published: 24th February 2021

The IEC 60601-1-2 standard is the international standard for testing medical equipment to EMC. The latest update adds a new test as more transmitter products are present in homes, offices (locations where ME equipment may be used), and hospitals. Three frequencies that are in use are listed: 30 kHz, 134.2 kHz, and 13.56 MHz. The standard IEC 61000-4-39 for testing fields in close proximity is referenced for this testing. To produce these fields can be very difficult.

Requirements IEC 60601-1-2:

Requirements IEC 61000-4-39:

The magnetic loops are specified in IEC 61000-4-39 and are readily available. However, producing a field through the loops might not be straight forward. The loop impedance changes over its frequency range, and the amplifier being used will not match this impedance. The result would be a much higher power requirement. This is very tricky to calculate and model before purchasing the correct power before you have the setup. In this case, we are not covering a frequency range but individual frequency points. This allows a matching network to be created to match the loop to the amplifier's impedance.

Basic setup:

Schlöder has created a system to match this requirement. The system is computer-controlled to set the field level and perform testing with report generation.

Required Equipment:

Generator and measurement:

CDG 7000-75-10 - 10 kHz – 250 MHz, 75 watt, amplifier. RF Generator, RF Meter x3 (monitored level, forward& reverse power). The system includes all cabling and HELIA 7 software required to meet the standard's requirements.

Accessories 9 kHz – 150 kHz:

  • MGA RL 120 – Radiating loop 120 mm as specified in IEC 61000-4-39 for 9kHz-150 kHz, IEC / EN 60601 - 1 - 2 and MIL-STD-461 / RS101, 3m cable
  • MGA LS 040 – Loop sensor 40 mm that attaches to MGA RL 120 at the correct distance of 50mm
  • MGA RL 120 NW – Matching Network for MGA RL 120, matches loop to 50 Ohms of the CDG 7000 amplifier acc. to IEC 61000-4-39 and IEC / EN 60601 - 1 - 2 for Immunity to magnetic fields 9kHz - 150 kHz

Accessories 150 kHz – 26 MHz:

  • MGA RL-100-3 – Radiating loop as specified in IEC 61000-4-39 and IEC / EN 60601 - 1 - 2 for 150 kHz - 26 kHz
  • MGA LS-040-1 – Loop sensor 40 mm that attaches to MGA RL 100-3 at the correct distance of 50 mm as specified in IEC 61000-4-39 and IEC / EN 60601 - 1 - 2
  • MGA RL 100-3 NW-60601 – Matching Network 60601 for MGA RL 100-3, matches MGA RL 100-3 to the IEC 60601-1-2 requirements
  • MGA RL 100-3 NW-61000 – Matching Network 61000 for MGA RL 100-3, matches MGA RL 100-3 to the IEC 61000-4-39 requirements

Optional:

MGA RL 100-3 Stand - Stand for MGA RL-100-3 for tests in more intensive magnetic fields

Added system advantages:

  • EUT monitoring: with Digital TTL signals & analog 0-10 VDC (automate threshold and report)
  • EUT Fail port: Stop test or mark report
  • SCPI interface control for universal support
  • Conducted immunity testing to IEC 61000-4-6, ISO, MIL, and other standards with the use of additional accessories CDNs, EMC clamps, BCI Clamps
category.title

Schlöder CDG 7000 Conducted Disturbances Test System

Generates Interference as Defined in IEC / EN 61000-4-6

The CDG 7000 is a new test generator for all interference immunity standards against conducted Interference induced by high frequency fields - including BCI tests (ISO 11452-4).

Find out more about Schlöder CDG 7000 Conducted Disturbances Test System

You Might also like

EMI Emissions Turntables

IEC 60601-1-2 ED. 4.1 / IEC 61000-4-39 New Requirement

The IEC 60601-1-2 standard is the international standard for testing medical equipment to EMC.

Read more
Equipment designed for the task

IEC 60601-1-2 ED. 4.1 New requirement, immunity to proximity magnetic fields based on IEC 61000-4-39

The IEC 60601-1-2 standard is the international standard for testing medical equipment to EMC. The latest update adds a new test as more transmitter products are present in homes, offices (locations where ME equipment may be used), and hospitals.

Read more
25 year old SESD 200 successfully calibrated

25 year old SESD 200 successfully calibrated

At the beginning of calendar year 2020, an old model of the "SESD 200" series successfully passed Schlöder's factory calibration in accordance with the IEC 61000-4-2 standard.

Read more

Sign up for the MCS Newsletter

You will receive all the latest test & measurement news and rental offers.